The Development of an External Quality Assessment System Using Inter-Laboratory Comparison for Real-Time PCR-Based Malaria Diagnosis in Regional Reference Laboratories under the Department of Disease Control, Thailand
Main Article Content
Abstract
Malaria remains a major global public health challenge, particularly in countries pursuing malaria elimination. Real-time polymerase chain reaction (Real-time PCR) is a highly sensitive and specific diagnostic method capable of detecting low-density and asymptomatic infections. However, a standardized External Quality Assessment (EQA) system for malaria diagnosis using Real-time PCR has not previously been established for the network of regional reference laboratories under Thailand’s Department of Disease Control (DDC). This study aimed to develop an External Quality Assessment system using an Inter-laboratory Comparison (ILC) approach and to evaluate the performance of regional reference laboratories for malaria diagnosis by Real-time PCR.
A descriptive study was conducted between September 2022 and October 2025 involving 10 regional reference laboratories under the DDC. Dried blood spot (DBS) panels were developed from residual malaria-positive blood specimens representing five human malaria species (Plasmodium falciparum, P. vivax, P. malariae, P. ovale, and P. knowlesi) and negative controls. The DBS panels were evaluated for homogeneity and stability before being distributed for two rounds of ILC assessment. Participating laboratories performed malaria detection using their routine Real-time PCR protocols and reported species identification and cycle threshold (Ct) values. Laboratory performance was assessed using concordance rates, Z-scores, and diagnostic performance indicators.
The developed DBS panels demonstrated acceptable homogeneity and stability, with Ct coefficient of variation values below 5% and Ct variation within ±1.0 cycle. Across two ILC rounds, the participating laboratories achieved a mean performance score of 99% (SD = 3.16). All laboratories obtained 100% accuracy in the first round, while nine laboratories achieved 100% accuracy in the second round and one laboratory obtained 80% due to a single discrepant result involving a low-concentration P. knowlesi specimen. Overall agreement among laboratories was excellent, with a concordance coefficient (Fleiss' Kappa) of 0.976 (p < 0.001), indicating almost perfect agreement. Most laboratories showed acceptable Z-scores ( ≤ 2), and Ct deviations remained within the predefined acceptance limits. Diagnostic performance analysis demonstrated 100% sensitivity, 95.24% specificity, 98.75% positive predictive value, and 100% negative predictive value.
The ILC-based EQA system developed in this study is suitable for evaluatingthe performance of malaria Real-time PCR laboratories. The findings demonstrate a high level of inter-laboratory agreement and confirm the capability of Thailand’s regional reference laboratory network to provide reliable molecular diagnosis for malaria surveillance and elimination activities. This system also supports compliance with ISO 15189 requirements and strengthens confidence in laboratory data used for malaria control and elimination programs in Thailand.
Downloads
Article Details

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
The opinions and content expressed in the articles are solely those of the individual authors and do not necessarily reflect the views of the journal or the authors’ affiliated institutions. Authors are fully responsible for the accuracy and correctness of their work.
Any reuse, reproduction, or redistribution of the articles, content, data, or images in other forms requires prior written permission from the Division of Vector Borne Diseases.
References
World Health Organization. World malaria report 2024 [Internet]. Geneva: World Health Organization; 2024 [cited 2026 Feb 13]. Available from: https://www.who.int/ publications/i/item/9789240104440
Department of Disease Control, Ministry of Public Health. Thailand malaria elimination strategy 2017–2026 [Internet]. Bangkok: Department of Disease Control; 2017 [cited 2026 Feb 13]. Available from: https://ddc.moph.go.th/dvb/pagecontent.php?page=22&dept=dvb (in Thai)
Wang B, Han SS, Cho C, Han JH, Cheng Y, Lee SK, et al. Comparison of microscopy, nested-PCR, and real-time PCR assays using high-throughput screening of pooled samples for diagnosis of malaria in asymptomatic carriers from areas of endemicity in Myanmar. J Clin Microbiol. 2014;52(6):1838-45. doi: 10.1128/JCM.03615-13.
Deo DA, Herningtyas EH, Intansari US, Perdana TM, Murhandarwati EH, Soesatyo MHNE. Difference between microscopic and PCR examination result for malaria diagnosis and treatment evaluation in Sumba Barat Daya, Indonesia. Trop Med Infect Dis. 2022;7(8):153. doi: 10.3390/tropicalmed7080153.
Mpina M, Stabler TC, Schindler T, Raso J, Deal A, Acuche Pupu L, et al. Diagnostic performance and comparison of ultrasensitive and conventional rapid diagnostic test, thick blood smear and quantitative PCR for detection of low-density Plasmodium falciparum infections during a controlled human malaria infection study in Equatorial Guinea. Malar J. 2022;21(1):99. doi: 10.1186/s12936-022-04103-y.
Lou Y, Shan Z, Chen Q, Kang F. Intra-laboratory and inter-laboratory variations analysis for HbA1c assays in China: using internal quality control and external quality assessment data. J Clin Lab Anal. 2025;39(8). doi: 10.1002/jcla.70036.
International Organization for Standardization. ISO 15189:2022 Medical laboratories—Requirements for quality
and competence [Internet]. Geneva: International Organization for Standardization; 2022 [cited 2026 Feb 24]. Available from: https://www.iacld.com/UpFiles/Documents/2e096ce5-485b-4f22-b7be-e557fb7d06f8.pdf
World Health Organization. Guidelines for malaria molecular testing [Internet]. Geneva: World Health Organization; 2022 [cited 2026 Feb 13]. Available from: https://iris.who.int/server/api/core/bitstreams/06f3d49f-cdce-4a2b-9ad9-a66395f40845/content
Moradian A, Goonatilleke E, Lin TT, Hatten-Beck M, Emrick M, Schepmoes AA, et al. Interlaboratory comparison of antibody-free LC-MS/MS measurements of C-peptide and insulin. Clin Chem. 2024;70(6):855-64. doi: 10.1093/clinchem/hvae034.
International Organization for Standardization. ISO 17043:2023 Conformity assessment—General requirements for proficiency testing [Internet]. Geneva: International Organization for Standardization; 2023 [cited 2026 Mar 8]. Available from: https://cdn.standards.iteh.ai/samples/80864/6038a69dfd204c3eb89f1fe3a6aac556/ISO-IEC-17043-2023.pdf
Thomson RM, Cunningham JA, Gatton MM, Murphy SC, Ade MdlP, Ding XC, et al. WHO malaria nucleic acid amplification test external quality assessment scheme: results of eleven distributions over 6 years. Malar J. 2025;24:94. doi: 10.1186/s12936-025-05282-0.
Cunningham J, Thomson R, Murphy S, Ade MdlP, Ding XC, Incardona S, et al. WHO malaria nucleic acid amplification test external quality assessment scheme: results of distribution programmes one to three. Malar J. 2020;19(1):129. doi: 10.1186/s12936-020-03200-0.
Rougemont M, Van Saanen M, Sahli R, Hinrikson HP, Bille J, Jaton K. Detection of four Plasmodium species in blood from humans by 18S rRNA gene subunit-based and species-specific real-time PCR assays. J Clin Microbiol. 2004;42(12):5636-43. doi: 10.1128/JCM.42.12.5636-5643.2004.
Lee PC, Chong ETJ, Anderios F, Lim YA, Chew CH, Chua KH. Molecular detection of human Plasmodium species in Sabah using PlasmoNex™ multiplex PCR and hydrolysis probe real-time PCR. Malar J. 2015;14:28. doi: 10.1186/s12936-015-0542-5.
Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159-74. doi: 10.2307/2529310.